Advanced software features on the HART transmitters (Rosemount 3102 and 3105 only)
Learn routine : (false echo registration)- The transmitter can learn to ignore up to four false echoes, caused by the pulse signal reflecting off obstructions
Empty tank mapping : When a tank is empty, the transmitter can learn to ignore up to four false echoes, without the need for user interaction
Present depth :The bottom reference can be automatically set using a known user-entered depth
Set as empty :When the tank is empty, the bottom reference can be automatically reset to the measured distance
Distance offset :The distance to the surface can be adjusted by a user-entered positive or negative offset value
Level offset :The level can be adjusted by a user-entered positive or negative offset value
Bottom blanking :The transmitter can be set to ignore an area of the tank bottom to avoid false echoes from obstructions